Air Sample Pump : PCXR4
The Universal PCXR4 personal or area air sampling pump features a 5 to 5000 ml/min flow range, timer, flow fault, and up to 12-hour run times for sampling particulates and gases/vapors. The pump is ideal for use with sorbent tubes, filter cassettes, size-selective samplers, sample bags, and impingers.
The Universal PCXR4 air sampling pump is UL Listed for intrinsic safety.
Pump Flow Rate | 5 to 5000 ml/min |
PC Compatible | No |
Programable | No |
Intrinsic Safety | Yes |
Battery Type (Power) | NiMH |
Approx Run Time | 12 |
Operating Range | 1000 to 5000 ml/min (UL Listed model) (5 to 500 ml/min requires the 224-26 Series adjustable tube holder) Requires 1/4-inch ID tubing |
Weight | 34 oz (964 gm) |
Dimensions | 5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm) |
Compensation Range | 1000 to 2500 ml/min at 40 inches water back pressure 3000 ml/min at 35 inches water back pressure 4000 ml/min at 20 inches water back pressure 5000 ml/min at 10 inches water back pressure |
Flow Control | Holds constant flow to ± 5% of set-point |
Run Time : Battery | NiCad Battery: 8 hrs minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used. See Table 3 NiMH Battery: 12 hrs minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used. See Table 2. Battery Eliminator: Pump provides extended runs. |
Run Time: Mains Adapter | 9999 minutes (6.8 days). Pump will shut off as run time cannot exceed timer display. See Timer Display Range below. |
Tubing | Requires 1/4-inch ID tubing |
Timer Display Range | 1 to 9999 minutes (6.8 days). Pump will shut off at 9999 minutes. To reset, restart the pump. |
Flow Indicator | Built-in rotameter with 250-ml division; scale marked at 1, 2, 3, 4, and 5 L/min |
Charging Time | 6 to 8.5 hrs with PowerFlex charger |
Power Supply | 6-V plug-in NiMH battery pack, rechargeable, 3.5-Ah capacity or 6-V plug-in NiCad battery pack, rechargeable, 2.0-Ah capacity A battery eliminator is available (see Ordering Information); use voids the UL Listing for intrinsic safety. |
Intrinsic Safety | UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C ATEX-approved model is available. Contact SKC. |
Operating Temp. | 32 to 104 °F (0 to 40 °C) |
Storage Temp. | -4 to 113 °F (-20 to 45 °C) |
Charging Temp. | 50 to 113 °F (10 to 45 °C) |
Operating Humidity | 0 to 95% non-condensing |
Flow Fault | If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters flow fault. During flow fault, the pump stops, displays FLOW FAULT, pauses timing functions, and displays elapsed time. Auto-restart is attempted up to 5 times. |
Low Battery Fault | Pump stops, displays LO BATT, pauses timing functions, and displays elapsed time. |
Battery Test | LCD indicator verifies battery condition prior to sampling. |
Time Display | LCD displays up to 9999 minutes (6.8 days) at which point the display rolls over to 0. |
Timing Accuracy | ± 0.05% (± 45 sec/day) |
Sampling Pause (Hold) | Allows user to temporarily halt air sampling without loss of timing data. Restart does not require resetting time. |
Multiple-tube Sampling | Built-in constant pressure regulator allows user to take up to four simultaneous samples with rates up to 500 ml/min each (total combined flow 1350 ml/min maximum) using the optional adjustable tube holder accessory |
RFI/EMI Shielding | Complies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1 Frequency range of the radiated susceptibility test was 27 to 1000 MHz CE marked |