Mavolux 5032 C BASE
This precision luxmeter
Due to its class C accuracy, this meter is used primarily as an industrial measuring instrument for planning, installing, inspecting and monitoring lighting systems, as well as for the assurance of compliance with specified lighting conditions.
Its four measuring ranges with automatic or manual range selection cover a broad span from 0.1 to 199,900 lx with an accuracy level of ± 3% ± 1 digit. V(λ) matching deviation, which amounts to f1‘ < 7.5%, is considerably better than the permissible error limit for class C specified in the standards.
At GOSSEN, great emphasis is placed upon reliability by means of calibration. For purposes of substantiation, a factory calibration certificate or a DAkkS calibration certificate can be ordered along. Depending on how the meter is used, we recommend a calibration interval of 12 to 24 months.
Thechnical Data | |
Illuminance [Lux] | 0.1 lx … 199 900 lx / 0.01 fc … 19990 fc |
Measuring ranges | 4 |
Measuring range selection | Automatic / manual |
Measuring rate | 2 per second |
Measuring method | Distance measurement |
Measuring sensor | Silicon photodiode with V (λ) filter |
Measurement cable | 1.5 m, firmly connected |
Measured value memory | 100 measured values |
Error limit – V(λ) adapted (f1‘), typical | < 7.5 % |
Error limit – overall error, typical | ≤ 15 % |
Accuracy | ± 3 % of reading ± 1 digit |
Battery | 2 x 1.5V mignon, type AA |
Automatic shutdown | 4 min / continuous operation |
Battery service life | Approx. 45 h with alkaline manganese battery |
Display | 3 1/2 digit LCD |
Operating elements | 6 keys |
Operating temperature | 0 °C to 50 °C |
Dimensions | 65 mm x 120 mm x 19 mm (meter) |
31 mm x 105 mm x 30 mm (probe) | |
Weight | 200 g without battery |
Delivery contents | Battery, operating instructions, calibration protocol |