AirChek TOUCH : Air Sampling Pump
The SKC AirChek TOUCH air sampling pump is the first personal sample pump with color touch screen navigation. Ultra-quiet and tough AirChek TOUCH provides extended flows from 5 to 5000 ml/min and high back pressure compensation for all your IH applications. Programming continuous, timed, and intermittent sampling is easy — no instruction manual needed. A powerful Li-Ion battery provides long run times. Charging, calibration, and PC communication are made easy with all-in-one cradles. Use AirChek TOUCH with DataTrac Pro Software for easy pump programming and advanced reporting.
The AirChek TOUCH air sampling pump is UL Listed for intrinsic safety.
Pump Flow Rate | 5 to 5000 ml/min |
PC Compatible | Yes |
Programable | Yes |
Intrinsic Safety | Yes |
Battery Type (Power) | Li-lon |
Approx Run Time | 20 |
Flow Range | Constant flow from 1000 to 5000 ml/min (low flow from 5 to 500 ml/min requires holder) |
Compensation Range | 5000 ml/min at 20 inches water back pressure 4000 ml/min at 30 inches water back pressure 3000 ml/min at 40 inches water back pressure 2000 ml/min at 50 inches water back pressure 1000 ml/min at 50 inches water back pressure |
Flow Control System | Isothermal, corrects for changes in back pressure, temperature, and atmospheric pressure |
Power | Removable rechargeable Lithium-ion (Li-Ion) battery, 7.4 V 2.6 Ah, 19.2 Wh or AC using charging cradle |
Flow Control | Holds constant flow to ± 5% of set-point after calibration |
Run Time | 20 hours at 2000 ml/min† 10 hours at 5000 ml/min† Indefinite run from charging cradle |
Accuracy | Flow control: ± 5% of set-point after calibration to desired flow Atmospheric pressure: ± 0.3 in Hg Temperature: ± 1 C |
Temperature Ranges | Operating: 32 to 104 °F (0 to 40 °C) Charging: 32 to 113 °F (0 to 45 °C) Storage: -4 to 113 °F (-20 to 45 °C) |
Humidity Ranges | Operating: < 95% RH, non-condensing Storage: < 95% RH, non-condensing |
Altitude | Corrects flow for changes in temperature (32 to 104 F/ 0 to 40 C) and ambient pressure up to 15,000 feet (4,572 meters) above and down to 4,500 feet (1,372 meters) below sea level |
Display/parameters | Color LCD/real-time flow rate, ambient temperature, ambient pressure, accumulated volume, elapsed time |
User Interface | Resistive touch screen with auto-dim and locking options |
Sound Level | Average 51.7 dB at 3-ft (1-m) distance using a 37-mm 0.8-µm MCE filter cassette |
Dimensions | 4.1 x 3.7 x 2.8 in (10.4 x 9.4 x 7.1 cm) |
Weight | 19.4 oz (550 gm) |
Certifications | • Intrinsic safety: UL class I, Div. 1, Groups A, B, C, and D; Class II, Div 1, Groups E, F, and G; Class III, Div. 1, hazardous locations when used with SKC battery pack model P75718; T-code T3C. Exia; Class I, Zone 0 Gp IIC (SKC Cat. No. 220-5000TC) • RoHS compliant • CE Marked |
Case Material | Polycarbonate with rubberized anti-static overmolding |
Charging Time | Approximately 3 hours |