890ZV SOFC Test Load

Product Descriptions
The bench-top Model 890ZV consists of a programmable electronic load, fuel and temperature controls, and data acquisition functions. A built-in power source makes the 890ZV particularly well suited for small single-cell SOFC work for research, laboratory, and educational use.

The 890ZV is optimized for cost and size, and offers extra hardware features and Scribner Associates’ FuelCell® software.

The 890ZV is typically used with a SOFC fuel cell, a suitable furnace, and a fuel management unit. The 890 Series is supported by the Scribner 885 Integrated Fuel Cell Potentiostat.

• Electronic load with two current ranges: 5 / 25 A, 10 V, 50 W
• Ramping cell/furnace temperature control
• Control signals for two (2) main mass flow controllers plus five (5) additional mass flow controllers
• Continuous real-time cell resistance and IR-free voltage measurement by current interrupt
• Whole cell voltage plus two high-impedance reference electrode inputs for half-cell data
• Optional internal impedance analyzer for EIS and continuous real-time high frequency resistance (HFR)
• Cell main terminals and sense inputs tolerant of non-isolated cell
• Three (3) internal controllers for anode and cathode humidifier and cell temperatures
• Current, voltage or power control mode
• Contact inputs for three pressure sensors or alarms
• Low voltage output signal to control purge gas valves and to indicate alarm condition
• RS485 digital interface for supported external devices
• The 890ZV requires a custom or third-party fuel management unit to make a complete test station.
• Safety features include detection of alarm conditions and automatic hardware shutdown for safe, reliable operation

 Electronic Load
 Maximum Load Current (2 ranges) 5 A, 25 A
 Maximum Load Voltage 10 V
 Maximum Load Power 50 W
 Minimum Load Resistance 3.3 V power supply used to allow zero load voltage at full current with up to 3.2 V drop in cell wiring
 Current Resolution 1 mA
 Current Accuracy 0.3% of full scale current of selected range
 Voltage Measurement and Data Acquisition
 Maximum Whole Cell Voltage  10.000 V
 Maximum Reference Electrode Voltage  9.999 V
 Whole Cell Sense Voltage Input Resistance  35 kΩ
 Reference Electrode Input Resistance  109Ω
 Voltage Resolution  1 mV
 Voltage Accuracy  +3 mV +0.3% of reading
 Voltage and Current Data Update Rate  100 Hz
 Fuel Interface
 Outputs for anode, cathode flow controllers  Two, Analog (0-5 V)
 Outputs for reformate flow controllers  Five, Analog (0-5 V) (optional)
 Alarm Inputs  Six: Three for gas pressures, three auxiliary
 Alarm Outputs  One, 5 V logic
 Fuel Solenoid Control  One, 5 V output (external relay needed, included with interface box)
 Temperature Controllers
 Quantity  Three (Anode, Cathode and Cell / Furnace)
 Type  On/off 5 V output (external SSR required)
 Set and Report Accuracy  +0.25% of span, +1 least significant digit
 Sensor Type  Thermocouple, Type K Standard, T or S optional
 Environment
 Operating Temperature  0-40 °C ambient; all specs given for 25 °C ambient
 Power Source  100-240 VAC, 50/60 Hz (auto select)
 Size  3U std rack mount, 19” W x 5.25″ H x 21″ D

885 Potentiostat

881 FRA for EIS & HFR

Back Pressure Units

850 Auto Multigas Unit

892e Data Expansion

Fuel Cell Fixture

Gas Mixing Interface

Heated Cuffs

850 Stand-Alone Humidifier

DI Water Tank

หากท่านสนใจกรุณากรอกรายละเอียด

Related Searches:

TESTA 1000 Solar Radiation chamber

highlyprecise and reproducible conditionsfor climatic and temperaturetesting in many industries.

Titrator : TitroLine 7500 KF trace

a compact, easy-to-use coulometric titrator for water determination according to Karl Fischer (KF) with a typical operating range of 1 ppm – 5% water.

TESTA BT – Battery Testing Chambers

battery testing solutions in line with the EUCAR hazard levels, ensuring safe, reliable, and sophisticated testing for energy storage systems.

Titrator : TITRONIC 300

perfect motor-driven piston burette for dosing tasks and manual titration

Chat with us on LINE

เราใช้คุกกี้เพื่อพัฒนาประสิทธิภาพ และประสบการณ์ที่ดีในการใช้เว็บไซต์ของคุณ คุณสามารถศึกษารายละเอียดได้ที่ นโยบายความเป็นส่วนตัว และสามารถจัดการความเป็นส่วนตัวเองได้ของคุณได้เองโดยคลิกที่ ตั้งค่า

Privacy Preferences

คุณสามารถเลือกการตั้งค่าคุกกี้โดยเปิด/ปิด คุกกี้ในแต่ละประเภทได้ตามความต้องการ ยกเว้น คุกกี้ที่จำเป็น

Allow All
Manage Consent Preferences
  • คุกกี้ที่จำเป็น
    Always Active

    ประเภทของคุกกี้มีความจำเป็นสำหรับการทำงานของเว็บไซต์ เพื่อให้คุณสามารถใช้ได้อย่างเป็นปกติ และเข้าชมเว็บไซต์ คุณไม่สามารถปิดการทำงานของคุกกี้นี้ในระบบเว็บไซต์ของเราได้

  • คุกกี้เพื่อการวิเคราะห์

    คุกกี้ประเภทนี้จะทำการเก็บข้อมูลการใช้งานเว็บไซต์ของคุณ เพื่อเป็นประโยชน์ในการวัดผล ปรับปรุง และพัฒนาประสบการณ์ที่ดีในการใช้งานเว็บไซต์ ถ้าหากท่านไม่ยินยอมให้เราใช้คุกกี้นี้ เราจะไม่สามารถวัดผล ปรังปรุงและพัฒนาเว็บไซต์ได้

Save