855 SOFC Test System
Product Descriptions
The 855 SOFC Test system is an integrated, turn-key Solid Oxide Fuel Cell Test System. This unit has a zero volt capability for lower measurement range. The addition of a programmable Power Supply extends the 855’s capabilities to include electrolysis mode. Automated, hands-free switching between SOFC and SOEC mode via FuelCell® software. The 855 SOFC system is known for it’s high performance and accuracy.
FuelCell® Software and Experimental Methods Manual are included. Optional Impedance Spectroscopy & HFR.
• Suitable for Button Cells and Planar Cells
• Multi Range Electronic load 5/25/50 A, 100 W, 20 V
• FuelCell® software for user-friendly computer controlled cell operation & experimentation
• Stainless Steel Anode humidifier with Bypass Valve and auto waterfill
• Programmable Power Supply for SOEC mode operation – 100 A / 5 V
• Power Supply mode includes Voltage & Current control with Constant, Scanning & Step-Stair experiments
• Up to five mass flow controllers available
• Includes Furnace Control hardware with redundant over-temperature protection
• Constant or stoichiometric-controlled reactant flow rate
• Zero Volt Capability
• Current, voltage or power control mode
• Continuous real time cell resistance and IR-free voltage measurement by Current Interrupt
• Whole cell voltage plus two high-impedance reference inputs for half-cell data
• Cell main terminals & sense inputs tolerant of non-isolated cell
• Safety features include detection of alarm conditions and automatic hardware shutdown for safe, reliable operation
Electronic Load: | |
Maximum Load Current | 5/25/50 A |
Maximum Load Power | 100 W |
Minimum Load Resistance | < 2 mΩ (100 mV @ 50 A at load terminals) |
Current Resolution | 1 mA for 5/25/50 A |
Current Accuracy | 0.3 % of full scale current rating |
Voltage Measurement and Data Acquisition | |
Maximum Whole Cell Voltage | 20 V |
Maximum Reference Electrode Voltage | 9.999 V |
Sense Lead Input Resistance | > 35 kΩ |
Voltage Resolution | I mV |
Voltage Accuracy | ±3 mV ±0.3 % of reading |
Voltage and Current Data Update Rate | 100 Hz |
Reactant Gas Control System | |
Mass Flow Control | Anode up to 1 SLPM and cathode up to 3 SLPM*. |
Software controlled mass flow controllers. Automatic N2 purge valves on anode and cathode | |
Temperature Controllers | |
Quantity | (3) Furnace, anode humidifier, cell temp |
Set and Report Accuracy | ±0.25 % of span, ±1 least significant digit |
Sensor Type | Thermocouple, type K (R/S also available) |
Temperature Limit Controller | (1) Furnace over-temp |
Humidifier | Anode only, sparger-type, single passivated 316L W heaters per bottle |
Temperature Range | Ambient to 99 ° C |
Environment | |
Operating Temperature | 0 – 40 °C |
Power Source | 120 V, 50/60 Hz, 900 W |
Enclosure Type | Single bench top enclosure |
Size and Weight | 18” H x 11” W x 19” D, 50 lb |
46 cm x 28 cm x 48 cm, 23 kg |
Scribner 855 SOFC Installation Video
885 Potentiostat
881 FRA for EIS & HFR
Back Pressure Units
850 Auto Multigas Unit
892e Data Expansion
Fuel Cell Fixture
Gas Mixing Interface
Heated Cuffs
850 Stand-Alone Humidifier