840 Fuel Cell Test System

Product Descriptions
The 840 Advanced Fuel Cell Test System is a turn-key test station for fuel cell research and university education programs. The 840 combines a computer-controlled instrument with fuel handling hardware in an integrated bench-top unit. The 840 is designed for use with small to large single cells or short stacks.

The Model 840 is targeted for applications requiring gas flow rates up to 12 SLPM and is based on the time-tested 850 series of turn-key fuel cell test stations. The 840 features dual range anode and cathode mass flow controllers for wide range flow control which is ideal for RH cycling. Also included are input selector valves and a 16-channel data acquisition module for additional voltage and temperature measurements.

The Model 840 uses Scribner’s 890 Multi-Range Electronic Load for accurate, wide-range current measurement. And, the Model 840 comes with FuelCell®, the industry’s most powerful and user-friendly application software for controlling station operation and experimentation.

• Ideal for large area single cell cells & short stacks
• 890 Electronic Load: 500W (12/62/125A or 25/125/250A) or 1 kW (50/250/500A), 20 V
• 892 Data Acquisition Module: 16 channels of voltage/temperature measurement integrated with FuelCell® software
• Dual Anode & Cathode mass flow controllers for enhanced accuracy over wide flow range
• Gas Selector Valves for Automated Switching between up to 3 Anode and 3 Cathode Gases
• Automated humidifier by-pass for wet/dry cycling
• Large capacity Anode & Cathode SS humidifiers with automatic water fill
• Flexible SS temperature controlled heated gas transfer lines
• FuelCell® software for user-friendly computer controlled cell operation & experimentation
• Constant or stoichiometric-controlled reactant flow rate
• Current, voltage or power control modes
• Continuous real time cell resistance and IR-free voltage measurement by Current Interrupt
• Whole cell voltage plus two high-impedance reference inputs for half-cell data
• Impedance spectroscopy & HFR option
• Safety features include detection of alarm conditions and automatic hardware shutdown for safe, reliable operation

 Electronic Load:
 Maximum Load Curren  5/25/50 A; 10/50/100 A; 12/62/125 A; 25/125/250 A (config. dependent)
 Maximum Load Power  125 W or 500 W (configuration dependent)
 Minimum Load Resistance  < 2 mΩ (100 mV @ 50 A at load terminals)
 Current Resolution  1 mA at low currents – up to 100 mA (current setting dependent)
 Current Accuracy  ±0.3% of full scale current of selected range
 Voltage Measurement and Data Acquisition:
 Max. Whole Cell Voltage  20 V Max
 Reference Electrode Voltage  9.999 V
 Voltage Resolution  1 mV
 Voltage Accuracy  ±3 mV ±0.3% of reading
 Voltage & Current Data Update Rate  100 Hz
 Whole Cell Sense Input Resistance  > 35 kΩ
 Reference Electrode Input Resistance  > 109 Ω
 Impedance Analyzer (Optional 880):
 Internal Impedance Analyzer Type  Single sine, one generator and two gain/phase measurement channels
 Internal Analyzer Frequency Range  1 mHz to 10 kHz
 Measurement Channels  Three: whole cell plus two half cell vs. Reference Electrode
 Reactant Gas Control System  All 316 SS construction (humidifiers, flow path, valves and mass flow controllers)
 with Swagelok® fittings Temperature controlled reactant transfer lines
 Mass Flow Control  Dual, software controlled mass flow controllers per channel, Anode: 6 SLPM (1 + 5 SLPM),
 Cathode: 12 SLPM (2 + 10 SLPM). Other sizes available on request. Automatic N2 purge valves
 Backpressure Control (Optional)  Manual or Automatic: 0 – 3 atm (0 – 30 PSIG).
 High Capacity: forced air condensers with large tanks and SS regulators.
 Temperature Controllers  Five: cell, anode humidifier, anode line, cathode humidifier,
 cathode line Set & Report Accuracy: ±0.25% of span,
 ±1 least significant digit Sensor Type: Thermocouple,
 Type T for cell (Type K optional for high temperature)
 Humidifiers  Dual sparger-type, passivated 316L, 1650 W heaters per bottle
 Temperature Range  Ambient to 99 °C Fill Method: Automatic water fill.
 Requires 3 atm (45 PSIG) minimum water feed or 1.4 atm (20 PSIG) above back pressure
 Environment:
 Operating Temperature  5 to 35 °C
 Power Source  220-240 V, 50-60 Hz, 20 A
 Enclosure Type:
 Single Bench Top Enclosure Size and Weight  91 cm H x 61 cm W x 61 cm D (+ 41 cm for heated gas lines),
 55 kg 36” H x 24” W x 24” D (+ 16” for heated gas lines), 120 lb.

885 Potentiostat

881 FRA for EIS & HFR

Back Pressure Units

850 Auto Multigas Unit

892e Data Expansion

Fuel Cell Fixture

Gas Mixing Interface

Heated Cuffs

850 Stand-Alone Humidifier

DI Water Tank

หากท่านสนใจกรุณากรอกรายละเอียด

Related Searches:

TESTA 1000 ตู้สำหรับจำลองการทดสอบการแผ่รังสีจากดวงอาทิตย์

วิเคราะห์และศึกษาข้อมูลเชิงลึกของผลิตภัณฑ์ที่สนใจภายใต้แสงแดด อุณหภูมิและความร้อนต่างๆ ที่สามารถควบคุมได้

Titrator : TitroLine 7500 KF trace

a compact, easy-to-use coulometric titrator for water determination according to Karl Fischer (KF) with a typical operating range of 1 ppm – 5% water.

TESTA BT – Battery Testing Chambers

battery testing solutions in line with the EUCAR hazard levels, ensuring safe, reliable, and sophisticated testing for energy storage systems.

Titrator : TITRONIC 300

perfect motor-driven piston burette for dosing tasks and manual titration

Chat with us on LINE

เราใช้คุกกี้เพื่อพัฒนาประสิทธิภาพ และประสบการณ์ที่ดีในการใช้เว็บไซต์ของคุณ คุณสามารถศึกษารายละเอียดได้ที่ นโยบายความเป็นส่วนตัว และสามารถจัดการความเป็นส่วนตัวเองได้ของคุณได้เองโดยคลิกที่ ตั้งค่า

Privacy Preferences

คุณสามารถเลือกการตั้งค่าคุกกี้โดยเปิด/ปิด คุกกี้ในแต่ละประเภทได้ตามความต้องการ ยกเว้น คุกกี้ที่จำเป็น

Allow All
Manage Consent Preferences
  • คุกกี้ที่จำเป็น
    Always Active

    ประเภทของคุกกี้มีความจำเป็นสำหรับการทำงานของเว็บไซต์ เพื่อให้คุณสามารถใช้ได้อย่างเป็นปกติ และเข้าชมเว็บไซต์ คุณไม่สามารถปิดการทำงานของคุกกี้นี้ในระบบเว็บไซต์ของเราได้

  • คุกกี้เพื่อการวิเคราะห์

    คุกกี้ประเภทนี้จะทำการเก็บข้อมูลการใช้งานเว็บไซต์ของคุณ เพื่อเป็นประโยชน์ในการวัดผล ปรับปรุง และพัฒนาประสบการณ์ที่ดีในการใช้งานเว็บไซต์ ถ้าหากท่านไม่ยินยอมให้เราใช้คุกกี้นี้ เราจะไม่สามารถวัดผล ปรังปรุงและพัฒนาเว็บไซต์ได้

Save