740 Membrane Test System

Product Descriptions
The 740 MTS is ideally suited for development of new ionomers and solid electrolytes. It offers rapid and accurate measurement of the through-thickness membrane resistance and conductivity as a function of temperature, humidity level and pressure.
The 740 MTS eliminates the time-consuming approach of catalyzing a membrane and assembling a fuel cell to evaluate the ionomer performance.

• Custom MTS4 Software for complete measurement process control and data acquisition
• CorrWare®, ZPlot® , MultiStat®, ZPlot·Lab™ & ZView® software for measurement and analysis of sample impedance
• Wet and dry gas mixing for accurate and precise RH control and rapid RH cycling
• Test chamber backpressure control, condenser and collection tank for high-temperature operation
• In-situ sample temperature and dew point measurement for real-time RH monitoring
• Uses bare (non-catalyzed) membranes
• Small sample size 6 mm x 30 mm
• Spring -loaded sample compression—easy, repeatable sample loading up to 400 PSI (2.76 MPa)
• Control and monitoring of multiple process temperatures
• Rapid, easy cell assembly with accurate and repeatable sample compression
• Cell lock-feature for safe operation at elevated temperature, pressure and H2 gas
• Automatic shutdown and N2 purge on alarm condition
• Detailed cell assembly instructions and data analysis procedure
• Native USB interface to PC for easy set-up
• Safety features include alarm conditions and automatic hardware shutdown for safe, reliable operation.

 Gas and Humidification System:
 Humidifier  316 SS material, designed for 100% gas saturation, auto water fill
 Mass Flow Control  Two; Wet and Dry, 0 – 500 sccm each (allows variable wet gas % and RH control)
 Sample Gas Selector  Selects N2 or H2/other
 Gas Pressure Switches  Two; N2 and H2/other
 Gas Dewpoint Meter  One, capacitance-type sensor in sample chamber
 Measurable Dewpoint  0 to 100 °C
 Set and Report Accuracy  +0.25% of span, +1 least significant digit
 Set and Report Resolution  0.1 0 C
 Sensor Type  Thermocouples, Type T
 Temperature (max)  180 °C (chamber, sample), 120 °C (humidifier), 130 °C (gas transfer line)
 Sample Chamber:
 Open/Close Mechanism  Threaded, screw-in cell head with integrated electrodes and locking-pin
 Electrode Clamp Mechanism  Mechanical, spring-loaded, integrated with cell head
 Electrodes  Platinum, proprietary four-terminal design (U.S. Patent No. 7,652,479)
 Temperature Range  Ambient to 180 °C
 Pressure Range  Ambient to 30 PSIG (207 kPa)
 Post-Chamber Gas Flow Path  Heat exchanger, condensate collection tank, precision manual back pressure regulator, vent port
 Physical and Environment:
 Operating Temperature  15 – 35 0C
 Power Source  100 – 120 VAC; 50/60 Hz
 Size, cm (in.)  46 H x 28 W x 48 D (18 x 11 x 19); excluding installed cell head
 Software inputs for each experiment  Chamber, humidifier and gas transfer line temperature set point; Pre-treatment (equilibration) time;
 Total gas flow rate, Wet gas %, Gas type (H2other or N2), and Back pressure (manually set)
 Number of steps per experiment  Up to 100

หากท่านสนใจกรุณากรอกรายละเอียด

Related Searches:

TESTA 1000 ตู้สำหรับจำลองการทดสอบการแผ่รังสีจากดวงอาทิตย์

วิเคราะห์และศึกษาข้อมูลเชิงลึกของผลิตภัณฑ์ที่สนใจภายใต้แสงแดด อุณหภูมิและความร้อนต่างๆ ที่สามารถควบคุมได้

Titrator : TitroLine 7500 KF trace

a compact, easy-to-use coulometric titrator for water determination according to Karl Fischer (KF) with a typical operating range of 1 ppm – 5% water.

TESTA BT – Battery Testing Chambers

battery testing solutions in line with the EUCAR hazard levels, ensuring safe, reliable, and sophisticated testing for energy storage systems.

Titrator : TITRONIC 300

perfect motor-driven piston burette for dosing tasks and manual titration

Chat with us on LINE

เราใช้คุกกี้เพื่อพัฒนาประสิทธิภาพ และประสบการณ์ที่ดีในการใช้เว็บไซต์ของคุณ คุณสามารถศึกษารายละเอียดได้ที่ นโยบายความเป็นส่วนตัว และสามารถจัดการความเป็นส่วนตัวเองได้ของคุณได้เองโดยคลิกที่ ตั้งค่า

Privacy Preferences

คุณสามารถเลือกการตั้งค่าคุกกี้โดยเปิด/ปิด คุกกี้ในแต่ละประเภทได้ตามความต้องการ ยกเว้น คุกกี้ที่จำเป็น

Allow All
Manage Consent Preferences
  • คุกกี้ที่จำเป็น
    Always Active

    ประเภทของคุกกี้มีความจำเป็นสำหรับการทำงานของเว็บไซต์ เพื่อให้คุณสามารถใช้ได้อย่างเป็นปกติ และเข้าชมเว็บไซต์ คุณไม่สามารถปิดการทำงานของคุกกี้นี้ในระบบเว็บไซต์ของเราได้

  • คุกกี้เพื่อการวิเคราะห์

    คุกกี้ประเภทนี้จะทำการเก็บข้อมูลการใช้งานเว็บไซต์ของคุณ เพื่อเป็นประโยชน์ในการวัดผล ปรับปรุง และพัฒนาประสบการณ์ที่ดีในการใช้งานเว็บไซต์ ถ้าหากท่านไม่ยินยอมให้เราใช้คุกกี้นี้ เราจะไม่สามารถวัดผล ปรังปรุงและพัฒนาเว็บไซต์ได้

Save