600 PEM Water Electrolysis Test System
600 Electrolyzer Test System (ETS), Advanced Test System for Research & Development of PEM Electrolyzers
The 600 ETS is a state-of-the-art, full-featured bench-top instrument for detailed evaluation and characterization of PEM electrolyzers. The 600 ETS includes an integrated Power Supply, Potentiostat, Impedance Analyzer for EIS & HFR, and real-time sensors for Product Flow Rate and Cross-over Monitoring. The 600 ETS is ideal for labs requiring detailed control, diagnostic and analysis of a single electrolzyer cell.
Powerful FlowCell-ETS® software offers full control and monitoring of the test cell. Built-in experiments include constant, scanning or step-stair voltage or current controlled measurements. Easy, hands-free switching between control of the voltage or current by the Potentiostat or Power Supply.
• Bench-top, fully integrated instrument for electrolyzer R&D
• 4 current range potentiostat (± 0.07 / 0.7 / 7 / 20 A, ±5 V) for accurate wide dynamic range current measurement
• Software controlled programmable Power Supply for operation up to 100 A / 5 V
• Automated switching between Potentiostat and Power Supply mode
• EIS and HFR (1 mHz to 10 kHz) in Potentiostat mode
• FlowCell-ETS® application software for complete system control, experiment sequencing, graphing and data acquisition
• EIS data compatible with ZView®, the world’s leading impedance analysis & equivalent circuit modelling software
• Integrated back pressure to 2 barg
• N2 purge on Cathode with Supplemental N2 on Anode
• Anode feed water recycling or once-thru operating mode
• In-line high-temperature Ion Exchange cartridge to maintain Anode feed water purity in recycle mode
• Includes safety features: E-Stop, Over Voltage/Current, Over Temperature or Gas Detection shutdown
• Integrated 892e Data Expansion Module for additional 8 temperature + 8 analog inputs, e.g., pressure transducers
• In-line, real-time O2 and H2 product flow rate monitoring
Cell Connections | 4-terminal (I+, I-, V+, V-) & differential Aux (REF) |
Cell Potentiostat: | |
Full Scale Current Ranges | ±20 / 7 / 0.7 / 0.07 A |
Current Resolution | 0.007% of range |
Current Limit of Error | ±1.0% of range |
Set and Read Voltage vs . WE | > ±5.000 V |
Cell Voltage Sense Lead | Differential w/ driven shields |
Voltage Measurement Resolution | 152 μV |
Sense Lead Input Resistance | 1.0 GΩ |
Modes of Operation | Constant, Scan, Step-Stair; V and I |
Cell Power Supply: | |
Maximum Current | 100 A |
Voltage Range | 0 – 5.000 V |
Power | Up to 500 W |
Impedance Measurement (Potentiostat Mode): | |
Frequency Range | 1 mHz to 10 kHz |
Measurement Types | Sweep EIS and single-frequency HFR real-time measurement, |
Whole Cell and Aux | |
Cell and Electrolyte Handling: | |
Flow Path | All 316 SS |
Feed Water Reservoir | 1 L, 316 SS, auto-water fill, conductivity probe |
Feed Water Supply Pump | 50 – 400 mL/min, software controlled |
Feed Water Temperature Range | Ambient to 95 oC |
Ion exchanger / Deionizer | In-line water recycling loop |
Back Pressure | Dual, 0-2 barg (0-30 PSIG), Manual |
Cell Temperature Range | Ambient to 120 oC |
Purge Gas | PC-controlled MFC on Negative; Manual on Positive |
Water/Gas Separator/Dehumidifiers | 2 (Negative, Positive), condensers & collection tanks |
Product Mass Flow Meters | 2 (1 SLM H2, 0.5 SLM O2) |
Additional Data Acquisition (892e): | |
Data Acquisition (892e) | 8 Temperature + 8 Analog (e.g., 0-5 V, 4-20 mA) |
Options: | |
In-Line Gas Sensors | H2 Transmitter for monitoring of H2 in Product O2 |
O2 Transmitter for monitoring of O2 in Product H2 | |
Physical and Environment: | |
Operating Temperature | 5-35 oC |
Power Source | 100-120 or 220-240 VAC, 50/60 Hz |
Size (excluding tubing connections) | 53 cm x 53 cm x 90 cm |
Weight | 64 kg (140 lb.) |