580 Battery Test System
Product Descriptions
The 580 Battery Test System was designed to meet the specific demands of our loyal customers.
This system is used to test both single cell batteries as well as super-capacitors.
Our powerful, flexible, user friendly BCycle® software is included with the instrument.
The 580 is used to test single cell batteries as well as super-capacitors.
• 6 current ranges – 1 A, 100 mA, 10 mA, 1 mA, 100 μA, 10 μA (Auto)
• 5 terminal measurement
• Control modes: current, voltage, power
• HFR for internal resistance
• 100 pts/sec sample rate
• BCycle® and BView® software – user-friendly, powerful, flexible
Voltage | |
Set Voltage | -2.000 V to +10.000 V |
Compliance Voltage | -2.000 V to +10.000 V |
Maximum Current | ±1.000 A |
Current Ranges | 1 A, 100 mA, 10 mA, 1 mA, 100 μA, 10 μA (Auto) |
Mode | Constant Current, Voltage or Power (charge, discharge) |
Control Accuracy | |
Voltage | < 0.1% of Positive Full Scale (at 25 °C) |
Current | < 0.1% of Full Scale of Range (at 25 °C) |
Measurement Accuracy | |
Voltage | < 0.05% of Positive Full Scale (at 25 °C) |
Current | < 0.1% of Full Scale of Range (at 25 °C) |
Overload Tolerance | |
Current | 110% of rating |
Voltage | 110% of rating |
Connection | 2, 3, 4 or 5-wire (I+, I-, V+, V-, AUX+) |
Sampling Rate | Max. 100Sa/s per channel (with 1 580-unit per computer) |
Impedance | 2 frequency HFR; user-defined frequencies |
Host Interferance | USB, full-speed, HID-compliant |
Channels | 8, independent, not electrically isolated from each other or USB interface |
Dimensions | 43 cm W x 22.5 cm H x 54 cm D (17 in. x 8.8 in. x 21.3 in.) |
Weight | 9 kg (20 lbs.) |
Environment | 5 to 35 ° C |
Power | 100-240 V, 50/60 Hz |