เครื่องกำเนิดอุณหภูมิ PTB 600

Product Descriptions
The PTB 600 Dry Block temperature calibrator offers industry-leading performance in an easy to use portable package with a wider range from 33 to 650 ºC. It is ideal for the calibration of Thermocouples and platinum resistance thermometers.  It has been designed for fast heating and cooling for convenient use in field.
The PTB 600 can also be used for Calibrating Infrared Sensors and Surface Temperature Sensors by using the optional inserts.
The new PTB 600 comes in two different models: TS and TC
• PTB600-TS Temperature Calibrator
• PTB600-TC Temperature calibrator with inputs for reference sensor and sensor-under-test
Fast and accurate calibration
Although featuring a large calibration volume of 35 x 185 mm,  the PTB600 heating block technology allow for fast heating and reduced time to stability. This translate in reduced cost of calibration, less downtime and less man power. All in all it saves money.
Interchangeable multi-holes inserts
The PTB600 uses interchangeable multi-holes inserts. The standard range of inserts accommodate most of the standard temperature sensors. Custom inserts are available on request to accommodate non standard temperature sensor testing.

  Specifications
  Temperature Range
  Ambient Temperature @ 23ºC   33°C to 650 °C
  Ambient Temperature @ 0ºC   10°C to 650°C
  Stability ( measured over 30 mins)   ± 0.05 °C
  Uniformity 
  Axial (lower 40 mm)    ± 0.35 °C
  Radial   ± 0.15 °C
  Accuracy
  with Internal Reference Sensor   ± 0.30 °C
  with External Reference Sensor SRS (*)   ± 0.15°C   ( TC version only)
  Display  Settings
  Set Point Range   -0°C to 650 °C
  Resolution   user-selectable 0.1 , 0.01 C
  Units   user-selectable C, F,K
  Calibration Volume
  Immersion depth   185 mm
  Insert Diameter   35 mm
  Standard Insert Supplied   Ø3.5-5-6.5-8.5-12.5 mm pockets
  Heating time   33 to 650 ºC : 36 min
  Cooling Time   650 to 100 ºC : 39 min
 100 to 50ºC : 19 min
  50 to 33ºC: 21 min
  Thermocouple Input (TC version)
  Type   K,N,R,S,J,E
  Range   0 to 1700 °C
  Accuracy   ± 1°C Include CJC error
  Resolution   user-selectable 0.1 , 0.01 °C
  RTD’S input ( TC version only)
  Type   Pt 100 (3/4 wires) IEC
  Range   -100 C to 700 °C
  Accuracy   ± 0.3 °C
  Resolution   user-selectable 0.1 , 0.01 °C
  General
  Computer interface   RS 232 as standard ( UBS adapter supplied as option) and EIUK software included
  Switch Test   Yes With automatic log of switch open and switch closed
  Ramp   Yes fully programmable
  Weight   10 Kg
  Dimensions   160x340x330 mm
  Power Requirement   115 o 230 V – 50/60 Hz
  Power   800 VA
  Notes   * Require SRS sensor and special UKAS matching calibration. Long term drift not included
  Scope of Supply
  Each PTB600 is supplied with: Power Cable, RS233 cable, switch test lead, standard insert, insert removal tool,
  traceable calibration certificate and instruction manual

หากท่านสนใจกรุณากรอกรายละเอียด

Related Searches:

TESTA 1000 ตู้สำหรับจำลองการทดสอบการแผ่รังสีจากดวงอาทิตย์

วิเคราะห์และศึกษาข้อมูลเชิงลึกของผลิตภัณฑ์ที่สนใจภายใต้แสงแดด อุณหภูมิและความร้อนต่างๆ ที่สามารถควบคุมได้

Titrator : TitroLine 7500 KF trace

a compact, easy-to-use coulometric titrator for water determination according to Karl Fischer (KF) with a typical operating range of 1 ppm – 5% water.

TESTA BT – Battery Testing Chambers

battery testing solutions in line with the EUCAR hazard levels, ensuring safe, reliable, and sophisticated testing for energy storage systems.

Titrator : TITRONIC 300

perfect motor-driven piston burette for dosing tasks and manual titration

Chat with us on LINE

เราใช้คุกกี้เพื่อพัฒนาประสิทธิภาพ และประสบการณ์ที่ดีในการใช้เว็บไซต์ของคุณ คุณสามารถศึกษารายละเอียดได้ที่ นโยบายความเป็นส่วนตัว และสามารถจัดการความเป็นส่วนตัวเองได้ของคุณได้เองโดยคลิกที่ ตั้งค่า

Privacy Preferences

คุณสามารถเลือกการตั้งค่าคุกกี้โดยเปิด/ปิด คุกกี้ในแต่ละประเภทได้ตามความต้องการ ยกเว้น คุกกี้ที่จำเป็น

Allow All
Manage Consent Preferences
  • คุกกี้ที่จำเป็น
    Always Active

    ประเภทของคุกกี้มีความจำเป็นสำหรับการทำงานของเว็บไซต์ เพื่อให้คุณสามารถใช้ได้อย่างเป็นปกติ และเข้าชมเว็บไซต์ คุณไม่สามารถปิดการทำงานของคุกกี้นี้ในระบบเว็บไซต์ของเราได้

  • คุกกี้เพื่อการวิเคราะห์

    คุกกี้ประเภทนี้จะทำการเก็บข้อมูลการใช้งานเว็บไซต์ของคุณ เพื่อเป็นประโยชน์ในการวัดผล ปรับปรุง และพัฒนาประสบการณ์ที่ดีในการใช้งานเว็บไซต์ ถ้าหากท่านไม่ยินยอมให้เราใช้คุกกี้นี้ เราจะไม่สามารถวัดผล ปรังปรุงและพัฒนาเว็บไซต์ได้

Save